Detecção de hipertensão ocular em Taco Pozo, Chaco, Argentina

Autores

  • Héctor S. Barni Centro de Ojos Quilmes, Quilmes, Buenos Aires, Argentina
  • Leila Galetto Hospital Salvador Mazza, Villa Angela, Chaco, Argentina
  • Sebastián Schlaien Centro de Ojos Quilmes, Quilmes, Buenos Aires, Argentina
  • Mariana F. Salom Hospital Salvador Mazza, Villa Angela, Chaco, Argentina. Sanatorio Central de Salud, Villa Angela, Chaco, Argentina
  • Celina Logioco Centro de Ojos Quilmes, Quilmes, Buenos Aires, Argentina
  • Virginia Zanutigh Centro de Ojos Quilmes, Quilmes, Buenos Aires, Argentina
  • Gerardo Valvecchia Centro de Ojos Quilmes, Quilmes, Buenos Aires, Argentina
  • Elena Barraquer Fundación Elena Barraquer, Barcelona, España

DOI:

https://doi.org/10.70313/2718.7446.v16.n04.259

Palavras-chave:

tonômetro de Goldmann, tonômetro iCarem, concordância, catarata, glaucoma

Resumo

Objetivos: Avaliar o grau de concordância em relação ao valor da pressão intraocular entre os tonômetros Goldmann e iCare em uma campanha de cirurgia de catarata.

Materiais e métodos: Estudo transversal no qual a PIO foi avaliada em pessoas que compareceram à cirurgia pré-operatória de catarata em um hospital rural em Taco Pozo (Chaco), Argentina, em agosto de 2023. Foi avaliado o grau de concordância entre os valores. PIO obtida com ambos os métodos de medição usando uma análise de Altman Bland, com um teste t pareado para indicar as diferenças entre os dois. O resultado da população geral e por subpopulações foi analisado em relação à PIO (maior ou menor que 21 mmHg).

Resultados: A PIO foi avaliada em 385 olhos. Os valores de PIO para a população geral foram (primeiro Goldmann e depois iCare) 15,6 ± 4,3 e 17,7 ± 6,0 (p<0,001); nos olhos com PIO até 21 mmHg foi de 14,9 ± 2,5 e 16,8 ± 3,9 (p<0,001); nos olhos com PIO > 21 mmHg foi de 27,4 ± 8,6 e 31,8 ± 11,5 (0,07).

Conclusão: Foi encontrado um bom nível de concordância entre a tonometria de Goldmann e o iCare para olhos com PIO > 21 mmHg. O iCare é considerado um método confiável para detectar hipertensão ocular em campanhas de catarata.

Downloads

Os dados de download ainda não estão disponíveis.

Referências

Torres RM. Los datos y la relevancia de estandarizar la práctica clínica, ¿una verdad de perogrullo? Oftalmol Clin Exp 2023;16: e1-e2.

Zeppieri M, Gurnani B. Applanation tonometry. En: StatPearls [en línea]. Treasure Island: StatPearls Publishing, June 11, 2023. Disponible en: https://pubmed.ncbi.nlm.nih.gov/35881737/

Mark HH. Armand Imbert, Adolf Fick, and their tonometry law. Eye (Lond) 2012; 26: 13-16.

Zakrzewska A, Wiącek MP, Machalińska A. Impact of corneal parameters on intraocular pressure measurements in different tonometry methods. Int J Ophthalmol 2019; 12: 1853-1858.

Cervino A. Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure. Br J Ophthalmol 2006; 90: 1444-1446.

Lambert SR, Melia M, Buffenn AN et al. Rebound tonometry in children: a report by the American Academy of Ophthalmology. Ophthalmology 2013; 120: e21-e27.

Esmael A, Ismail YM, Elhusseiny AM et al. Agreement profiles for rebound and applanation tonometry in normal and glaucomatous children. Eur J Ophthalmol 2019; 29: 379-385.

ElMallah MK, Asrani SG. New ways to measure intraocular pressure. Curr Opin Ophthalmol 2008; 19: 122-126.

Marini M, Da Pozzo S, Accardo A, Canziani T. Comparing applanation tonometry and rebound tonometry in glaucomatous and ocular hypertensive eyes. Eur J Ophthalmol 2011; 21: 258-263.

Gao F, Liu X, Zhao Q, Pan Y. Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer. Exp Ther Med 2017; 13: 1912-1916.

Rödter TH, Knippschild S, Baulig C, Krummenauer F. Meta-analysis of the concordance of Icare™ PRO-based rebound and Goldmann applanation tonometry in glaucoma patients. Eur J Ophthalmol 2020; 30: 245-252.

Ashano O, Oderinlo O, Ogunro A, Ashano E. A comparison of goldmann applanation tonometry and rebound tonometry measurements among patients attending a glaucoma clinic in Southwest Nigeria. Niger J Clin Pract 2022; 25: 1542-1547.

Bilgeç MD, Atalay E, Sözer Ö et al. The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes. Int Ophthalmol 2020; 40: 849-857.

Giavarina D. Understanding Bland Altman analysis. Biochem Med (Zagreb) 2015; 25: 141-151.

Publicado

2023-12-20

Edição

Secção

Artigos Originais

Como Citar

[1]
2023. Detecção de hipertensão ocular em Taco Pozo, Chaco, Argentina. Oftalmología Clínica y Experimental. 16, 4 (Dez. 2023), e357-e365. DOI:https://doi.org/10.70313/2718.7446.v16.n04.259.